Quickly Assessing Fluorescence Bleaching in LSM700 Microscopes with ZEN and Fiji

Protocol

Initial Settings

  • On the LSM700, make sure you have all settings (Laser, Gain, Pixel Size, Speed) ready
  • Do not use averaging or Z stacks

Prepare for acquisition

  • Using the "Regions" tool in ZEN, create a rectangle in a region of your image with signal
  • In the Regions menu, uncheck "Bleach", but keep "Acquisition" & "Analysis"
  • Check "Use bounding rectangle blablabla"
  • Check "Time Series"
  • Use 100 timepoints (or more if you want)

TODO, illustrate image better

Run Acquisition

  • Use Start Experiment
  • Switch to the "Mean ROI" tab on the left of the image

Copy the data and run the macro

  • On the table, right click and select "Copy Data"
  • Open Fiji and open the macro "Assess Bleaching.ijm" (Drag & Drop the file into Fiji)

  • Run the macro. It will use the data that is currently on the Windows clipboard

See the results


Message with number of frames before 50% of intensity is lost.
This is also reflected in a new line in a Table called Bleach Assessment
Finally, you can observe the plot to see if the exponential fit worked

Last Author
oburri
Last Edited
Mar 7 2019, 13:02

Event Timeline

oburri created this document.Mar 7 2019, 12:17
oburri edited the content of this document. (Show Details)
oburri edited the content of this document. (Show Details)Mar 7 2019, 13:01
oburri edited the content of this document. (Show Details)
oburri added subscribers: romainGuiet, chiarutt.
oburri edited the content of this document. (Show Details)Mar 8 2019, 15:07
chiarutt edited the content of this document. (Show Details)Apr 25 2019, 12:54
oburri edited the content of this document. (Show Details)Jan 25 2023, 10:59