Page MenuHomec4science

PerkinElmer Operetta FCS
Updated 656 Days AgoPublic

Tutorial

Outdated : F8105712

Worflow and Data management

Relocate Settings

SettingsRelocate Settings

Using Fiji to process your Operetta images

Mind the Byte

One full well from a 96-well plate (4.5mm x 4.5mm field of view)

Full resolution acquisition 20x air objective
1Gb per plane (Single channel, single slice, single timepoint)

Example acquisition, 2 channels, 11 slices : 9.35 GB per well

Write archive

It takes ~1 hour for 100 GB

Harmony 4.9 installation

Start Autorun.exe
\\svfas6\biop\biop\Informatics\Software Library\Harmony\2019-10-17 SW-Update Harmony 4.9 MDe
Select: Office PC > install

NOTE: If the computer states that "You need to restart the computer to install", restart it once and retry. If the problem persists, do the following:
  1. Run Regedit.exe
  2. Find the Windows Registry: HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Control\Session Manager : PendingFileRnameOperations
  3. Delete the PendingFileRnameOperations registry (right click > delete > ok)
  4. You can now proceed to do the installation

The Region and Language Settings of the PC should be in English UK or US otherwise the software crashes when you start it.

Connect the Operetta database

  1. Start Harmony 4.9
  2. Go to settingsdata managementdatabase settings
  3. Add a new database (click on +)
  4. Computer name: SV-01-366
  5. Alias name: Operetta CLS

Plate Reference

BrandReference NameRef NumberQuantityPriceComments
PerkinElmerPhenoPlate 96 (previously Cell Carrier Ultra)605530240CHF 365.00works with 20x et 63x water
PerkinElmerPhenoPlate 96 (previously Cell Carrier Ultra)6055308160EUR 1241.00
PerkinElmerViewPlate-96F600518250EUR 390.00Can't access the wells on the border with high mag. obj. + issue with finding focus using 40x & 63x
PerkinElmerViewPlate-96F60052255x10EUR 372.00
Falcon96Well ImagingPlateFA-3532194 x 8CHF 206-290Can't access the wells on the border with high mag. obj. + issue with finding focus using 63x
Cellvis12 Well Cellvis4925231x20EUR 265.00Works with 40x water and ?

Export flat-fielded images

It is possible to export the flat-fielded image by using the script located in \\svraw1\ptbiop\public\0-Software\Operetta\Export_Flatfielded_Images inside the Harmony Software

This is a small tool which can apply the flat-field correction. It is based on the image processing core of Harmony (acapella.exe). It runs on Harmony 4.8 and seems to work fine. Should you still have Harmony 4.6 you probably have to edit the batch file for the correct path to acapella.exe. Unfortunately I can’t test it on older versions, but I don’t really expect problems.

  • First you have to export measurements from Harmony. On the Operetta CLS this can even be exports with “Referenced Images”, i.e. you will just get an “index.ref.xml” file. Or if you prefer the normal export, the “index.idx.xml” is also fine.
  • Then you run the batch file, which just calls the acapella.exe interpreter with the script file as parameter. This script will then ask you for an index.xml-file. Select the file that was created during the Harmony export.
  • It will create a new folder and write all flatfield corrected images into that folder. It uses a similar naming convention as Harmony, just with the prefix “flex_”. It will also generate a new appropriate index.xml file, in case your system works with the xml files.

These scripts can run on any Harmony installation that has a valid license dongle.

Tips and Tricks

Digital Phase Contrast (DPC)

NOTE: When setting up the DPC parameters, find a field where there are no artifacts or floating dead cells as these will affect the optimization.
  1. Find the focus position (offset) in brightfield where you see your cells the least. That is, the most in-focus plane.
  2. Set your exposure time and laser power so that you are using a good part of the histogram (max values around 50000 gray levels
  3. Set yourself on "High Contrast" Mode and hit "Snapshot DPC"
  4. The system will acquire 50 images and find the ones with the highest contrast to use as top plane and lower plane
  5. Set the DPC to "Manual"
  6. Change the Filter and Speckle Scale in the following way": Speckle to 0 in order to keep mitotic cells, otherwise they get removed. Filter to 0.5 if your cells have holes in them, or larger than 1 if there is too much background around the cells.

Viewing fused wells

When acquiring images in Test mode, you can select all of the fields, Right Click and select "Wells and Fields, Realistic".
This opens a fused version of the images in a separate window.

Adding an overview image to the well grid.

After you acquire a test images of multiple tiles, Select the images and select "Use Background for Well"
The fused image will appear as the background now, to help you select fields of interest when setting up the Acquisition

Using 35mm Petri Dishes

If you want to image a petri dish with glass bottom (No 1.5 - 170 um thickness), you can use the adapted X-wing holder designed by Oli. The layout to select is petri dish 35mm.

Using ibidi slides

Please have a look at https://github.com/BIOP/tool-ibidi-slide-holder:

Autofocus Failure: Debugging

In case there is a problem with the autofocus, you can call the Cockpit Data Plotter to see how it is trying to make the autofocus on the system.

WINPerkinElmerServiceFieldApp
Password: Scan4FAS


This plot shows the result for a "Two Peak" autofocus (In this case for the 40x Water objective, so from left to right: Water-Glass interface followed by Glass-Water. The sample would be on the right of this plot.


Example of failed autofocus: The large peak on the right here represents the interface of a transwell that is inside the well. Because the system focuses on the "largest peaks" it is unable to find a similar second peak and thus fails.

Option 1:switch to the "One Peak" mode.

In the example above, it would be useful as then the system would use the transwell as a reference (Because it is the strongest peak)

Option 2: Change the geometry of the plate

If we change the geometry of the plate so that the system expects it to be lower than it actually is:


In this case, if we change the value of H from 1.00 to 0.50, the Operetta will just stop scanning for reflections before finding the transwell, resulting in a successful Two Peak detection.

In the case of the CellVis Plates, two configurations were made for the Transwells.


Use the Two Peak AF plate for standard plates without transwells

Flatfield and Orientation Matrix

These two calibrations are computed on the fly for each experiment In case there is no Flatfield correction available, this is because there were not enough images that Harmony considered of sufficient quality to estimate a flatfield. This can happen when images touch the borders of the wells. The software will inform you after the experiment is run


Top: Successful flatfield estimation from enough images (30 images in total)
Bottom: Unsuccessful flatfield,experiment only contained 4 images in total.

Last Author
chiarutt
Last Edited
Jul 20 2022, 11:12

Document Hierarchy

Event Timeline

oburri created this document.Nov 3 2017, 15:09
oburri edited the content of this document. (Show Details)
bozzo edited the content of this document. (Show Details)Jan 29 2018, 11:27
romainGuiet edited the content of this document. (Show Details)Feb 14 2018, 13:55
bozzo edited the content of this document. (Show Details)Apr 5 2018, 09:36
bozzo edited the content of this document. (Show Details)Apr 20 2018, 10:00
romainGuiet edited the content of this document. (Show Details)May 29 2018, 11:38
bozzo edited the content of this document. (Show Details)May 29 2018, 12:47
bozzo edited the content of this document. (Show Details)Jun 14 2018, 10:35
bozzo edited the content of this document. (Show Details)Jun 14 2018, 10:57
chiarutt edited the content of this document. (Show Details)Nov 6 2018, 12:20
romainGuiet edited the content of this document. (Show Details)Apr 8 2019, 13:17
romainGuiet edited the content of this document. (Show Details)Apr 15 2019, 17:04
chiarutt edited the content of this document. (Show Details)Oct 18 2019, 11:08
chiarutt edited the content of this document. (Show Details)Jun 16 2020, 14:12
romainGuiet edited the content of this document. (Show Details)Jul 28 2020, 11:07
oburri edited the content of this document. (Show Details)Oct 16 2020, 08:39
chiarutt edited the content of this document. (Show Details)Feb 22 2021, 19:01
chiarutt edited the content of this document. (Show Details)Jul 21 2021, 08:25
chiarutt edited the content of this document. (Show Details)Jul 21 2021, 08:30
oburri edited the content of this document. (Show Details)Aug 5 2021, 08:40
oburri edited the content of this document. (Show Details)Aug 5 2021, 09:08
chiarutt edited the content of this document. (Show Details)Sep 30 2021, 15:30
chiarutt edited the content of this document. (Show Details)Oct 4 2021, 17:24
oburri edited the content of this document. (Show Details)Oct 19 2021, 16:32
oburri edited the content of this document. (Show Details)Oct 19 2021, 16:36
oburri edited the content of this document. (Show Details)Oct 26 2021, 10:15
chiarutt edited the content of this document. (Show Details)Jan 18 2022, 14:24
oburri edited the content of this document. (Show Details)Jan 19 2022, 16:01
oburri edited the content of this document. (Show Details)Jan 19 2022, 16:07
chiarutt edited the content of this document. (Show Details)May 31 2022, 10:14
chiarutt edited the content of this document. (Show Details)May 31 2022, 10:28
chiarutt edited the content of this document. (Show Details)Jul 20 2022, 11:12